Patient Monitor User Manual EMC Information
- 195 -
B.3 Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
X8 X10 and X12 are intended for use in the electromagnetic environment specified below. The
customer or the user of X8 X10 and X12 should assure that they are used in such an
environment.
Immunity
test
IEC/EN 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Conducted
RF
IEC/EN
61000-4-6
Radiated RF
IEC/EN
61000-4-3
3 V
rms
150 kHz to 80 MHz
6Vrms
c
in ISM
bands between
0.15 MHz and 80
MHz
3 V/m
80 MHz to 2.7 GHz
See Table 1
3 V
rms
150 kHz to
80 MHz
6Vrms
c
in
ISM bands
between
0.15 MHz
and 80 MHz
3 V/m
80 MHz to
2.7 GHz
Comply with
Table 1
Portable and mobile RF
communications equipment should be
used no closer to any part of X8 X10
and X12, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
Pd 2.1
150 KHz to 80 MHz
Pd 2.1
80 MHz to 800 MHz
Pd 3.2
800 MHz to 2.7 GHz
E/6 Pd
at RF wireless
communications equipment bands
(Portable RF communications
equipment (including peripherals such
as antenna cables and external
antennas) should be used no closer than
30 cm (12 inches) to any part of the
monitor, including cables specified by
the manufacturer).
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter