Product Manual

Patient Monitor User Manual Networked Monitoring
- 28 -
fails to be connected, please try to restart the monitor and connect again.
4 Use the wireless device recommended by EDAN, otherwise some exceptional
situations such as frequent network disconnection may occur on the monitor.
5 The wireless driver is compatible with channels 1-11 only.
6 When signal intensity is level 2 or less, signal may be unstable and quality of the
signal transmission may be degraded.
7 The user can click Update button or re-enter the Wi-Fi List to select updated Wi-Fi
site.
8 When the monitor is connected to MFM-CMS/Gateway via the wireless network, the
user should set the router to a secure encryption/authentication and use the
non-dictionary password.
Recommended options: WPA/WPA2 Personal (supports AES/TKIP);
Other options: none or WPA/WPA2 Enterprise (includes TLS/TTLS /PEAP).
4.3 Network Disconnected Alarms
To configure the network disconnected alarms, select Menu > Maintenance > User Maintain >
Alarm Setup and choose Disconnect Alarm which can be set to On or Off. The alarm is off by
default.
NOTE:
1 When the monitor is connected with the central monitoring system, you must set
Disconnect Alarm to On.
2 If Disconnect Alarm occurs during audio alarm paused or audio alarm off status, the
monitor will prompt a sounding alarm with information of NetWork Disconnect.
During the network disconnected status, activating audio alarm paused or audio alarm
off function can disable the audio alarm signal of Disconnect Alarm.
4.4 Connecting the Monitor to MFM-CMS
The monitor can be connected to the central monitoring system (V2.65 and above). Through the
network:
1. The monitor sends patient information, real-time monitoring or measurement data to the
central monitoring system.
2. The real-time monitoring information is displayed on the central monitoring system as the
same to the monitor, and the central monitoring system can perform some bilateral control.
For example: changing patient information, receiving patient, discharging patient and so
forth.
For detailed information, please refer to MFM-CMS Central Monitoring System User Manual.
And the monitor supports HL 7 protocol.
NOTE:
1 Use wired instead of wireless networking when connecting the monitor to central