monitoring system in the operating room because the ESU will interfere with a
wireless network, which may cause networking failure.
2 Make sure the network connection between the monitor and MFM-CMS is in good
condition when the time synchronization function on the monitor is active. (Default
setting is on. Setting path: Menu > Maintenance > User Maintain > Date/Time
Setup > Sync Time). If the setting is on, the monitor will accept time synchronization
from MFM-CMS.
3 The time synchronization function might not be available to all software versions of
MFM-CMS. Consult our technical service department or your local distributor for
more information.
4 When deploying the network of the monitor and MFM-CMS, it is recommended to
isolate the network and the Intranet system of the hospital by using VLAN so as to
ensure the network security. Only trusted devices are allowed to join the VLAN
network.
4.5 HL7 Communication
The monitor supports HL7 protocol to upload data. Select Menu > User Maintain and input user
maintain password > Security. In this menu:
Set HL7 to On/Off. When it is set to On, set Firewall to On to protect against hacker
attacking. To avoid hacker attacking, setting HL7 to Off is normally recommended.
Set HL7 Encryption method to Off or TLS (default).
To set HL7 IP address of the client-side, select User Maintain > Network Maintain.
Click Import Certificate to install/upgrade the Certificate via USB flash drive. The certificate
issued by Certificate Authority (CA) is recommended and self-signed certificate should be
avoided. For detailed steps of importing certificates, please refers to service manual.
For more information about HL7 communication, refer to HL7 Communication Protocol Service
Manual.