Patient Monitor User Manual Alarms
- 33 -
corresponding parameter area.
WARNING
1 When the alarm is set to Off, the monitor won’t give an alarm prompt even if an alarm
occurs. In order to avoid endangering the patient’s life, the user should use this
function cautiously.
2 Prior to monitoring, make sure that the alarm limit settings are appropriate for your
patient.
3 Setting alarm limits to extreme values may cause the alarm system to become
ineffective. It is recommended to use the default settings.
4 In HR alarm limit setting process, the bottom will display ExtremeTachy or
ExtremeBrady threshold value that has been set. HR high alarm limit should be less
than or equal to ExtremeTachy threshold value, and HR low alarm limit should be
more than or equal to ExtremeBrady threshold value.
5.4.2 Audio Alarm Paused
You can temporarily prevent alarms from sounding by pressing the hardkey on the front
panel or pressing shortcut key on the screen.
You can set the alarm pause time as desired. The default alarm pause time is 120 s.
1. Select Menu > Maintenance > User Maintain, and enter the required password.
2. Select Alarm Setup, and set Pause Time to 60 s, 120 s, or 180 s.
When alarms are paused,
The audio alarm is turned off, and no alarms are sounding.
The visual alarm indications are still displayed.