Product Manual
Patient Monitor User Manual Alarm Information
- 38 -
Message
Cause
Alarm Level
Pauses/min
High
The measurement value of Pause/min is greater than
high alarm limit that has been set.
User-selectable
PVCs High
The measurement value of PVCs is greater than high
alarm limit that has been set.
User-selectable
VEB
The delayed ventricular beats detected in normal
heartbeats occur more than or equal to 2 times within 30 s.
User-selectable
Multiform
PVCs
Different forms of ventricular premature beats are
detected in 15 beats.
User-selectable
IPVC
The single ventricular premature beat between 2 sinus
beats with normal interval occurs more than or equal to
3 times within 30 s.
User-selectable
PAC Bigeminy
The dominant rhythm of N, A, N, A, N, A, and the
rhythm number exceeds the number of threshold value
that has been set (N = supraventricular beat, A = atrial
beat).
User-selectable
PAC Trigeminy
The dominant rhythm of N, N, A, N, N, A, N, N, A, and
the rhythm number exceeds the number of threshold
value that has been set.
User-selectable
Low
Voltage(Limb)
None of the signal amplitudes of I, II and III leads
exceeds that of the alarm threshold that has been set.
PS: this alarm is available for 5, 6 or 10 electrodes only,
not available for 3 electrodes.
User-selectable
RESP
RESP APNEA
RESP waveform cannot be detected within the set apnea
alarm delay time.
High
RR High
RR measuring value is above upper alarm limit.
User-selectable
RR Low
RR measuring value is below lower alarm limit.
User-selectable
SpO
2
SpO
2
High
SpO
2
measuring value is above upper alarm limit.
User-selectable
SpO
2
Low
SpO
2
measuring value is below lower alarm limit.
User-selectable
SpO
2
No Pulse
The signal of the measurement site is too weak due to
insufficient blood supply and environmental factors, so
the monitor can’t detect the pulse signal.
High
SpO
2
Desat
SpO
2
measuring value is below the SpO
2
Desat limit.
High
PR High
PR measuring value is above upper alarm limit.
User-selectable