Patient Monitor User Manual Alarm Information
- 42 -
Message
Cause
Alarm Level
Action Taken
ECG V2 Lead Off
ECG electrode V2 falls off
the skin or the ECG cable
V2 falls off.
Low
properly connected.
ECG V3 Lead Off
ECG electrode V3 falls off
the skin or the ECG cable
V3 falls off.
Low
ECG V4 Lead Off
ECG electrode V4 falls off
the skin or the ECG cable
V4 falls off.
Low
ECG V5 Lead Off
ECG electrode V5 falls off
the skin or the ECG cable
V5 falls off.
Low
ECG V6 Lead Off
ECG electrode V6 falls off
the skin or the ECG cable
V6 falls off.
Low
ECG Signal
Exceeded
ECG measuring signal is
beyond measuring range.
Low
Check lead
connection and
patient condition
ECG Noise
ECG measuring signal is
greatly interrupted.
Low
ECG Comm Fail
ECG module failure or
communication failure
High
Stop measuring
function of ECG
module, and notify
biomedical engineer
or manufacturer’s
service staff.
RESP
RESP Comm Fail
RESP module failure or
communication failure
High
Stop measuring
function of RESP
module, and notify
biomedical engineer
or the manufacturer’s
service staff.
RESP Noise
RR cannot be measured due
to patient movement.
Low
Check whether the
RESP leads are well
connected. Keep the
patient calm for
better monitoring.