Patient Monitor User Manual Alarm Information
- 43 -
Message
Cause
Alarm Level
Action Taken
RR Exceed
RR measuring value is out of
the measure range.
Medium
Check whether
interference to the
respiratory signal
exists. And check
whether the patient is
breathing normally;
breathing too rapidly
or too slowly may
endanger patients
life.
RESP Cardiac
Artifact
No RESP waveform can be
detected due to apnea or
shallow breathing of the
patient.
High
Check whether the
patient is breathing
normally. Take
measures to help the
patient breathe
normally when
necessary. If the
patient is breathing
normally, try to
adjust the electrode
position on the
patient in order to
reduce the
interference of
cardiogenic artifact.
SpO
2
SpO
2
Sensor Off
SpO
2
sensor may be
disconnected from the patient
or the monitor.
Low
Make sure the sensor
is well connected to
the patient’s finger
or other parts. Make
sure the monitor and
cables are well
connected.
SpO
2
Comm Fail
SpO
2
module failure or
communication failure
High
Stop using
measuring function
of SpO
2
module, and
notify biomedical
engineer or
manufacturer’s
service staff.