Product Manual
Patient Monitor User Manual Alarm Information
- 46 -
Message
Cause
Alarm Level
Action Taken
NIBP Loose Cuff
Cuff is not properly wrapped
or no cuff is connected.
Low
Properly wrap the
cuff.
NIBP Weak Signal
Cuff is too loose or patient
pulse is too weak.
Low
Use other methods to
measure blood
pressure.
NIBP Range
Exceeded
All of SYS, DIA and MAP
values are beyond the
measurement range.
High
SYS(NIBP)
Overrange
SYS (NIBP) value is beyond
the measurement range.
High
DIA(NIBP)
Overrange
DIA (NIBP) value is beyond
the measurement range.
High
MAP(NIBP)
Overrange
MAP (NIBP) value is
beyond the measurement
range.
High
NIBP Interference
Signal noise is too large or
pulse rate is not regular due
to the patient movement.
Low
Make sure that the
patient under
monitoring is
motionless.
NIBP Leak Test
Error
Fail to deflate normally
during the leak test, so NIBP
leak test cannot be finished.
Low
Test again. If the
problem still exists,
contact your service
personnel.
TEMP
TEMP T1 Sensor Off
Temperature cable of TEMP
channe1 may be disconnected
from the monitor.
Low
Make sure that the
cable is properly
connected
TEMP T2 Sensor Off
Temperature cable of TEMP
channe2 may be disconnected
from the monitor.
Low
Make sure that the
cable is properly
connected.
Excessive T1
TEMP1 measuring value is
beyond measuring range.
High
Check sensor
connection and
patient condition
Excessive T2
TEMP2 measuring value is
beyond measuring range.
High