Patient Monitor User Manual Alarm Information
- 47 -
Message
Cause
Alarm Level
Action Taken
TEMP Comm Fail
TEMP module failure or
communication failure.
High
Stop measuring
function of TEMP
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
T1 Calibration
Failed
T1 calibration failed.
High
Please check whether
the module works
properly.
T2 Calibration
Failed
T2 calibration failed
High
IBP
YY Sensor Off (YY
stands for the IBP
label name: Art, PA,
CVP, RAP, LAP, ICP,
P1 and P2)
IBP sensor falls off.
Medium
Check the sensor
connection and
reconnect the sensor.
IBP Catheter Off
IBP catheter falls off due to
patient movement.
High
Check the catheter
connection and
reconnect it.
IBP Sensor Err
Malfunction in the IBP
sensor or in the extension
cable.
Medium
Replace the IBP
sensor or the
extension cable.
YY Comm Fail (YY
stands for the IBP
label name: Art, PA,
CVP, RAP, LAP,
ICP, P1 and P2)
IBP module failure or
communication failure
High
Stop measuring
function of IBP
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
C.O.
C.O. TI Sensor Off
C.O. TI sensor not connected
Low
Insert injective
temperature sensor.
C.O. TB Sensor Off
C.O. TB sensor not
connected
Low
Insert TB sensor.