Patient Monitor User Manual Alarm Information
- 49 -
Message
Cause
Alarm Level
Action Taken
Battery Error
Battery failure or high
temperature protection
Low
Replace the battery
and restart the
monitor. Or suspend
charging until the
battery cools and
then continue
charging. If the
problem still exists,
contact your service
personnel.
Battery Current Too
High
The battery current is too
high.
Low
Stop using the
battery, contact your
service personnel.
Battery Charge
Voltage Too High
The battery charging voltage
is too high
Low
Recorder Out Of
Paper
Recorder Out Of Paper
Low
Please install the
paper
Recorder Probe
Overheated
The probe of recorder is
overheated.
Low
Stop recording and
retry after the probe
cools.
Printer Unavailable
The selected printer is not
available.
Low
Check whether the
network connection
is in good condition
and whether the
printer is
malfunctioning.
Insufficient storage
space
Less than 10 M space is left
in the storage device.
Low
Delete some data in
the removable device
or use another
removable device.
Read-only storage
device
The storage device is
read-only.
Low
Repair the removable
device or replace it
with a new one.
Storage device
damaged
Storage device is damaged.
Low