Patient Monitor User Manual Alarm Information
- 50 -
Message
Cause
Alarm Level
Action Taken
Network Disconnect
In distributed alarm system,
the monitor’s network is
disconnected.
Low
1) Check if the
network cable is well
connected.
2) Check if the
MFM-CMS is turned
on.
3) Check if the IP of
bedside monitor and
MFM-CMS are on
the same network
segment.
Audio Failed
Audio circuit connection is
abnormal, or loudspeaker
falls off.
High
Stop using the
monitor and notify
the manufacturer’s
service staff.
Network traffic
anomaly
Abnormal network traffic has
been detected. The data traffic
exceeds the limit.
High
Disconnect the
network to make the
monitor work
properly, and then
contact the
professionals
authorized by
manufacturer to
check the network
problem.
6.3 Prompts
Message
Cause
ECG ARR Learning
The QRS template building required for Arr. Analysis is in
process.
Unable to analyze ST
The ST algorithm cannot produce valid ST value, which
may be caused by the large change in the measured value of
connected cardiogram ST or ventricular pacing.
Unable to analyze QT
QT algorithm cannot generate valid QT for more than 10
minutes (or 1 minute during startup).
QT Baseline Overrange
After modifying the calculation formula, the QTc parameter
value exceeds the range.
Unable to analyze ECG
The arrhythmia algorithm cannot analyze ECG data reliably.