Patient Monitor User Manual Alarm Information
- 52 -
Message
Cause
Venipuncture Ending
Finish the assisting venipuncture and the cuff begins to
deflate.
NIBP Simul
NIBP Simul function is turned on.
Be sure the cuff is disconnected
from monitor
In Cleaning Mode, the user clicks the Start Cleaning
button.
Cleaning succeeded
Cleaning finished successfully.
Cleaning failed
Abnormal air pressure in cleaning mode.
Cleaning in progress
The monitor is in cleaning progress.
Place Probe On Measure Place
Probe isnt placed on the measurement site.
CO
2
Standby
Turn from measuring mode to standby mode, making the
module in energy-saving status.
Please Press 'Zero'.
Enter the IBP zeroing menu, and zeroing is not performed
yet.
Zero OK
IBP completes zeroing.
Pulsatile Pressure Zero Fail.
During the zeroing process, pressure fluctuation is
excessive.
Pressure out of normal
range,Fail.
During the zeroing process, pressure value is beyond the
zeroing range.
Sensor Off, Fail!
Perform zeroing when the sensor is off.
Invalid Time,Zero Fail.
Time is not set up prior zeroing.
Unable to Calibrate in Demo
Mode
Perform zeroing in Demo Mode.
Zeroing...
Zeroing is in progress.
Please Press 'Calibrate'.
Enter the Calibration menu, and Calibration is not
performed yet.
Calibration OK
Calibration is completed.
Pulse Pressure Calibration
Failed
During the Calibration process, pressure fluctuation is
excessive.
Pressure out of range
During the Calibration process, pressure value is beyond
the Calibration range.
Zeroing and Calibration
Failed
Zeroing is not performed prior calibration.
Sensor Off, Fail.
Perform calibration when the sensor is off.
Invalid Time, Calibration Fail.
Time is not set up prior calibration.