QTc and ΔQTc alarm limits are listed as follows: unit (ms)
Adjustable Range
QTc
200~800
ΔQTc
30~200
RESP alarm limits are listed as follows: unit (rpm)
Patient Type
Adjustable Range
ADU
6~120
PED/ NEO
6~150
SpO
2
alarm limits are listed as follows (unit %):
Adjustable Range
SpO
2
20~100
SpO
2
Desat Limits are listed as follows (unit %):
Adjustable Range
SpO
2
Desat Limit
20~99
NOTE:
User can set the range through User Maintain > Alarm Setup > SpO
2
Desat Limit,
SpO
2
Desat Limit should be less than or equal to SpO
2
alarm low limit.
PR
alarm limits are listed as follows: unit (bpm)
Adjustable Range
PR (SpO
2
)
30~300
PR (NIBP)
40~240
PR (IBP)
30~300
NIBP alarm limits are listed as follows: unit (mmHg)
Patient Type
Adjustable Range
ADU
SYS
DIA
MAP
25~290
10~250
15~260
PED
SYS
DIA
MAP
25~240
10~200
15~215
NEO
SYS
DIA
MAP
25~140
10~115
15~125