Patient Monitor User Manual Managing Patients
- 57 -
Height: Enter the patient’s height, with unit: cm or inch.
Weight: Enter the patient’s weight, with unit: kg or lb.
NOTE:
1 Creating new patient and updating patient will clear the history data in the monitor
associated with the patient.
2 For Bed No., user can select English, Chinese, Russian through switching keyboard
language, and select special characters through .
7.2.1 Patient Category and Paced Status
The patient category setting determines the algorithm which the monitor uses to process and
calculate some measurements, the safety limits that are applied for some measurements, and the
alarm limit ranges.
The paced setting determines whether the monitor shows pacemaker pulses or not. When Pace is
set to Off, pace pulses are filtered and therefore do not show in the ECG wave.
WARNING
1 Changing the patient category may change the arrhythmia and NIBP alarm limits.
Always check alarm limits to make sure that they are appropriate for your patient.
2 For paced patients, you must set Paced to On. If it is incorrectly set to Off, the monitor
could mistake a pace pulse for a QRS and fail to give an alarm during asystole.
7.3 Quick Admit
If you do not have the time or information to fully admit a patient, you can use Quick Admit to
quickly admit a patient and complete the rest of the patient information later. To quickly admit a
patient, please:
1. Select the shortcut key on the screen directly, or
2. Select Menu > Patient Setup > Quick Admit, then a message is displayed to ask the user to
confirm to update patient.
3. Click on No to cancel this operation; click on Yes to continue and the Quick Admit window
is displayed.
4. Configure Type and Pace to the correct setting and click Yes to finish the quick patient
admission operation. If you want to quit the operation, click No.
7.4 Barcode Admit
Barcode scanner can recognize patient information directly and quickly, which can provide
convenience and reduce mistakes for users.