Service manual

Chapter 5: Troubleshooting 37
DICOM
Table 5.2: DICOM Troubleshooting
Error Message Tiller Error Code Cause Troubleshooting
Socket
communication
failed
TSOCKET_CONNECT_FAILURE Invalid network
configuration.
Wrong port
number.
Application is not
running.
Printer is offline.
Using Ping, verify that the
Printer/Archiver is connected.
If Ping fails, check the devices IP
address, Edge IP address,
Subnet mask, and Gateway IP
address.
If Ping is OK, use Verify to check
if device is available.
If Verify fails:
a) Check the Printer/Archivers
Port configuration on the Edge.
b) Ensure that the Printer is
online and the Archivers
application is running.
Archiver
transaction
failed
TDICARCH_OPEN_FAILURE Wrong Capture
Type selected
Verify that the Archiver supports
the selected Capture Type setting,
e.g., US Image, SC Image or
US-Ret Image.
Printer
transaction
failed
TDICPRNT_OPEN_FAILURE Wrong Image
settings
Verify that the printer supports
the selected Image settings. E.g,.
Color (RGB) or Grayscale
(Monochrome)
DICOM network
communication
failed
TDNETWORK_OPEN_FAILURE Device does not
recognize Edge,
rejects association
Verify that Edge AE Title or IP
address is correctly configured on
the Printer/Archiver.
Note: Some devices require that the
Imaging modality (Edge) be
recognized in order to accept
images. This requires configuration
on the device.
Internal failure
detected
TDNETWORK_READ_FAILURE Invalid DICOM
Attribute
Check Edge Printer DICOM
settings for correctness (e.g., film
size, format)