Chapter 7: Maintenance 53
Chapter 7: Maintenance
This chapter contains information to help you properly care for the system, transducers, and accessories.
Periodic Maintenance
No periodic or preventive maintenance is required for the system, transducers, or accessories other than cleaning and
disinfecting the transducer after every use. (See “Cleaning and disinfecting transducers” on page 55.) There are no
internal adjustments or alignments required and there are no internal components that require periodic testing,
calibration, adjustment, or alignment. Performance tests are described in Chapter 8, “Performance Testing” of this
manual. Performing maintenance procedures not described in this manual may void the product warranty.
Local regulations may require electrical safety testing.
Contact SonoSite Technical Support for any maintenance questions. (See “Technical Support (USA, Canada)” on
page 1.)
Cleaning and disinfecting
Use the recommendations in this section when cleaning or disinfecting the ultrasound system, transducer, and
accessories. Use the cleaning recommendations in the peripheral manufacturer’s instructions when cleaning or
disinfecting peripherals.
For recommended cleaners and disinfectants, see the disinfectant list available on the CD included with the ultrasound
system and on www.sonosite.com.
WARN ING:
Disinfectants and cleaning methods listed are recommended by SonoSite for
compatibility with product materials, not for biological effectiveness. Refer to the
disinfectant label instructions for guidance on disinfection efficacy and
appropriate clinical uses.
WARN ING:
The level of disinfection required for a device is dictated by the type of tissue it
contacts during use. To avoid infection, ensure that the disinfectant type and the
solution strength and duration are appropriate for the equipment. For information,
see the disinfectant label instructions and the recommendations of the
Association for Professionals in Infection Control and Epidemiology (APIC) and
the FDA.
WARN ING:
To prevent contamination, the use of sterile transducer sheaths and sterile
coupling gel is recommended for clinical applications of an invasive or surgical
nature. Do not apply the transducer sheath and gel until you are ready to perform
the procedure.
Caution:
Some transducer sheaths contain natural rubber latex and talc, which can cause
allergic reactions in some individuals. Refer to 21 CFR 801.437, User labeling
for devices that contain natural rubber.