Troubleshooting
Cleaning and disinfecting
Use the recommendations in this section when
cleaning or disinfecting the ultrasound system,
transducer, and accessories. Use the cleaning
recommendations in the peripheral manufacturer’s
instructions when cleaning or disinfecting
peripherals.
For recommended cleaners and disinfectants, see
the disinfectant list available on www.sonosite.com.
Cleaning and disinfecting the
ultrasound system
The exterior surface of the ultrasound system and
the accessories can be cleaned and disinfected
using a recommended cleaner or disinfectant.
WARNING:
Disinfectants and cleaning methods
listed are recommended by
SonoSite for compatibility with
product materials, not for biological
effectiveness. Refer to the
disinfectant label instructions for
guidance on disinfection efficacy
and appropriate clinical uses.
WARNING:
The level of disinfection required
for a device is dictated by the type
of tissue it contacts during use. To
avoid infection, ensure that the
disinfectant type and the solution
strength and duration are
appropriate for the equipment. For
information, see the disinfectant
label instructions and the
recommendations of the
Association for Professionals in
Infection Control and Epidemiology
(APIC) and the FDA.
WARNING:
To prevent contamination, the use
of sterile transducer sheaths and
sterile coupling gel is
recommended for clinical
applications of an invasive or
surgical nature. Do not apply the
transducer sheath and gel until you
are ready to perform the
procedure.
Caution:
Some transducer sheaths contain
natural rubber latex and talc, which
can cause allergic reactions in
some individuals. Refer to 21 CFR
801.437, User labeling for devices
that contain natural rubber.
WARNING:
To avoid electrical shock, before
cleaning, disconnect the system
from the power supply or remove
from the mini-dock or docking
system.
WARNING:
To avoid infection always use
protective eyewear and gloves
when performing cleaning and
disinfecting procedures.
WARNING:
To avoid infection, ensure that the
solution expiration date has not
passed.
Caution:
Do not spray cleaners or
disinfectant directly on the system
surfaces. Doing so may cause
solution to leak into the system,
damaging the system and voiding
the warranty.
Caution:
Do not use strong solvents such as
thinner or benzene, or abrasive
cleansers, since these will damage
the exterior surfaces.
Caution:
Use only recommended cleaners
or disinfectants on system surfaces.
Immersion-type disinfectants are
not approved for use on system
surfaces.