User guide
Chapter 6:  Troubleshooting and Maintenance 93
Troubleshooting
To clean and disinfect a transducer (wipe 
method)
1 Disconnect the transducer from the system.
2 Remove any transducer sheath.
3 Clean the surface using a soft cloth lightly 
dampened in a mild soap or detergent cleaning 
solution to remove any particulate matter or 
body fluids.
Apply the solution to the cloth rather than the 
surface.
4 Rinse with water or wipe with water-dampened 
cloth; then wipe with a dry cloth.
5 Mix the disinfectant solution compatible with the 
transducer, following disinfectant label 
instructions for solution strengths and 
disinfectant contact duration.
6 Wipe surfaces with the disinfectant solution.
7 Air dry.
8 Examine the transducer and cable for damage 
such as cracks, splitting, or fluid leaks. 
If damage is evident, discontinue use of the 
transducer, and contact SonoSite or your local 
representative.
To clean and disinfect a transducer 
(immersion method)
1 Disconnect the transducer from the system.
2 Remove any transducer sheath.
3 Clean the surface using a soft cloth lightly 
dampened in a mild soap or compatible cleaning 
solution to remove any particulate matter or 
body fluids.
Apply the solution to the cloth rather than the 
surface.
4 Rinse with water or wipe with water-dampened 
cloth, and then wipe with a dry cloth.
5 Mix the disinfectant solution compatible with the 
transducer, following disinfectant label 
instructions for solution strengths and 
disinfectant contact duration.
6 Immerse the transducer into the disinfection 
solution not more than 12-18 inches 
(31-46 cm) from the point where the cable 
enters the connector.
Follow the instructions on the disinfectant label 
for the duration of the transducer immersion.
7 Using the instructions on the disinfectant label, 
rinse to the point of the previous immersion, and 
then air dry or towel dry with a clean cloth.
8 Examine the transducer and cable for damage 
such as cracks, splitting, or fluid leaks.
If damage is evident, discontinue use of the 
transducer, and contact SonoSite or your local 
representative.
Cleaning and disinfecting the battery 
To clean and disinfect a battery (wipe 
method)
1 Remove the battery from the system.
Caution:
Do not allow disinfectant to contact 
metal surfaces. Use a soft cloth 
lightly dampened in a mild soap or 
compatible cleaning solution to 
remove any disinfectant that 
remains on metal surfaces.
Caution:
Attempting to disinfect a 
transducer or transducer cable 
using a method other than the one 
included here can damage the 
transducer and void the warranty.
Caution:
To avoid damaging the battery, do 
not allow cleaning solution or 
disinfectant to come in contact 
with the battery terminals.