User guide
108
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The Edge ultrasound system is intended for use in the electromagnetic environment specified
below. The customer or the user of the Edge ultrasound system should assure that it is used in
such an environment.
Emissions Test Compliance Electromagnetic Environment
RF emissions
ClSPR 11
Group 1 The Edge ultrasound system uses RF energy only
for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
ClSPR 11
Class A The Edge ultrasound system is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network which supplies buildings used for
domestic purposes.±
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Edge ultrasound system is intended for use in the electromagnetic environment specified
below. The customer or the user of the Edge ultrasound system should assure that it is used in
such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
AC Power
±6.0KV contact
±8.0KV air
±6.0KV contact
±8.0KV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30%.