User guide

110
Power
Frequency
Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Conducted RF
IEC 61000-4-6
3Vrms
150 kHz to 80 MHz
3 Vrms Portable and mobile RF
communications equipment
should be used no closer to
any part of the Edge
ultrasound system including
cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended Separation
Distance
d
= 1.2
Radiated RF
IEC 61000-4-3
3Vim
80 MHz to 2.5 GHz
3 V/m
d
= 1.2
80 MHz to 800 MHz
d
= 2.3
800 MHz to 2,5 GHz
Where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in meters (m).
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity (continued)
The Edge ultrasound system is intended for use in the electromagnetic environment specified
below. The customer or the user of the Edge ultrasound system should assure that it is used in
such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment
P
P
P