User guide

Chapter 8: Acoustic Output 123
Acoustic Output
Output display
The system meets the AIUM output display standard for MI and TI (See“Related guidance
documents” on page 125). Table 3 indicates for each transducer and operating mode when
either the TI or MI is greater than or equal to a value of 1.0, thus requiring display.
Note: The D2x transducer has a static continuous wave (CW) output. This output is fixed.
Therefore, TI and MI values cannot be changed by any system controls available to the user.
Table 3: TI or MI 1.0
Transducer Model Index
2D/
MMode
CPD/
Color
PW
Doppler
CW
Doppler
C11x/8-5 MI No No No
TIC,TIB, or TIS No Yes Yes
C60x/5-2 MI Yes No No
TIC, TIB, or TIS No No Yes
D2x/2 MI ———No
TIC,TIB, or TIS ———Yes
HFL38x/13-6 MI Yes Yes Yes
TIC, TIB, or TIS No Yes Yes
HFL50x/15-6 MI Yes Yes Yes
TIC, TIB, or TIS No No Yes
ICTx/8-5 MI No No No
TIC, TIB, or TIS No No Yes
L25x/13-6 MI Yes No No
TIC, TIB, or TIS No No Yes
L38x/10-5 MI No Yes Yes
TIC, TIB, or TIS No Yes Yes
L38xi/10-5 MI Yes Yes Yes
TIC, TIB, or TIS No Yes Yes
P10x/8-4 MI No Yes Yes No
TIC, TIB, or TIS Yes Yes Yes Yes