User guide
Chapter 8: Acoustic Output 131
Acoustic Output
(a)This index is not required for this operating mode; value is <1.
(b)This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason
listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 8: Transducer Model:
C60x
Operating Mode:
2D
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value 1.0 (a) — — — (b)
Associated Acoustic
Parameter
p
r.3
(MPa) 1.69
W
0
(mW) #— —#
min of
[W
.3
(z
1
),I
TA.3
(z
1
)]
(mW)
—
z
1
(cm) —
z
bp
(cm) —
z
sp
(cm) 4.7 —
d
eq
(z
sp
) (cm) —
f
c
(MHz) 2.84 # — — — #
Dim of A
aprt
X(cm) #—— — #
Y(cm)
#—— — #
Other Information
PD (μsec) 0.579
PRF (Hz) 5440
p
r
@PII
max
(MPa) 2.679
d
eq
@Pll
max
(cm) —
Focal Length FL
x
(cm) #—— #
FL
y
(cm) #—— #
I
PA.3
@MI
max
(W/cm
2
)
197.7
Operating
Control
Conditions
Control 1: Exam Type Abd
Control 2: Optimization Res/
Gen
Control 3: Depth 11/
13 cm
Control 4: THI On
Control 5: MB (Multi Beam) On










