User guide
136
(a)This index is not required for this operating mode; value is <1.
(b)This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason
listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 13: Transducer Model:
HFL38x
Operating Mode:
CPD/Color
Index Label MI
TIS TIB
TIC
Scan
Non-scan
Non-
scan
A
aprt
1A
aprt
>1
Global Maximum Index Value 1.1 1.0 — — — (b)
Associated Acoustic
Parameter
p
r.3
(MPa) 2.556
W
0
(mW) 37.69 — —#
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW) —
z
1
(cm) —
z
bp
(cm) —
z
sp
(cm) 1.2 —
d
eq
(z
sp
)(cm) —
f
c
(MHz) 5.328 5.324 — — — #
Dim of A
aprt
X(cm) 0.44 — — — #
Y(cm)
0.4 — — — #
Other
Information
PD (μsec) 0.525
PRF (Hz) 2597
p
r
@PII
max
(MPa) 3.187
d
eq
@Pll
max
(cm) —
Focal Length FL
x
(cm) 1.32 — — #
FL
y
(cm) 2.5 — — #
I
PA.3
@MI
max
(W/cm
2
)
325.5
Operating
Control
Conditions
Control 1: Mode Color Color — — — —
Control 2: Exam Type Any Ven — — — —
Control 3: Optimization/Depth/PRF Low/
3.3 cm/
Any
Med/
2.7 cm/
2841
————
Control 4: Color Box Position/Size Any Top/
Short
————