User guide

146
(a)This index is not required for this operating mode; value is <1.
(b)This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason
listed. (Reference Global Maximum Index Value line.)
Data are not applicable for this transducer/mode.
Table 23: Transducer Model:
L38x
Operating Mode:
PW Doppler
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value 1.0 2.0 2.6 (b)
Associated Acoustic
Parameter
p
r.3
(MPa) 2.345
W
0
(mW) —84.94 84.94 #
min of
[W
.3
(z
1
),I
TA.3
(z
1
)]
(mW)
z
1
(cm)
z
bp
(cm)
z
sp
(cm) 0.8 1.3
d
eq
(z
sp
) (cm) 0.4685
f
c
(MHz) 5.01 5.05 5.05 #
Dim of A
aprt
X(cm) 1.80 1.80 #
Y(cm)
0.4 0.4 #
Other Information
PD (μsec) 1.29
PRF (Hz) 1008
p
r
@PII
max
(MPa) 2.693
d
eq
@Pll
max
(cm) 0.2533
Focal Length FL
x
(cm) —5.54 #
FL
y
(cm) —2.5 #
I
PA.3
@MI
max
(W/cm
2
)
284.5
Operating
Control
Conditions
Control 1: Exam Type Any Vas Vas
Control 2: Sample Volume 1mm 12mm 12mm
Control 3: PRF 1008 Any Any
Control 4: Sample Volume Position Zone 0
(top)
Zone 7 Zone 7