User guide

148
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#No data are reported for this operating condition since the global maximum index value is not reported for
the reason listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 25: Transducer Model:
L38xi/10-5
Operating Mode:
MMode
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value 1.5 (a) 1.2 (b)
Associated Acoustic
Parameter
p
r.3
(MPa)
3.54
W
0
(mW)
—# 37.1 #
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
z
1
(cm)
z
bp
(cm)
z
sp
(cm)
1.0
0.9
d
eq
(z
sp
)(cm)
0.49
f
c
(MHz)
5.76 # 5.20 #
Dim of A
aprt
X(cm) —# 1.86#
Y(cm)
—# 0.40#
Other Information
PD (μsec) 0.146
PRF (Hz) 1600
p
r
@PII
max
(MPa)
4.32
d
eq
@Pll
max
(cm)
0.49
Focal Length FL
x
(cm)
—# #
FL
y
(cm)
—# #
I
PA.3
@MI
max
(W/cm
2
)
776
Operating
Control
Conditions
Control 1: Exam Type Any Any
Control 2: Optimization Gen Pen
Control 3: Depth
4.7 cm
7.3 - 9.0
cm