User guide
154  
(a)This index is not required for this operating mode; value is <1.
(b)This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason 
listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 31: Transducer Model: 
P10x
Operating Mode: 
CW Doppler
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value (a) — (a) — 2.1 2.0
Associated Acoustic
Parameter
p
r.3
(MPa) #
W
0
(mW) —# 40.72 30.00
min of 
[W
.3
(z
1
),I
TA.3
(z
1
)]
(mW)
—
z
1
(cm) —
z
bp
(cm) —
z
sp
(cm) # 0.7
d
eq
(z
sp
) (cm) 0.36
f
c
(MHz) # — # — 4.00 4.00
Dim of A
aprt
X(cm) — # — 0.320 0.16
Y(cm)
—# — 0.70.7
Other Information
PD (μsec) #
PRF (Hz) #
p
r
@PII
max
(MPa) #
d
eq
@Pll
max
(cm) 0.27
Focal Length FL
x 
(cm) —# — 0.92
FL
y
 (cm) —# — 5.0
I
PA.3
@MI
max
(W/cm
2
)
#
Operating
Control 
Conditions
Control 1: Exam Type Card Card
Control 2: Depth Any Any
Control 3: Zone Zone 3 Zone 0