User guide
160
((a)This index is not required for this operating mode; value is <1.
(b)This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason
listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 37: Transducer Model:
SLAx
Operating Mode:
PW Doppler
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value (a) — (a) — 1.1 (b)
Associated Acoustic
Parameter
p
r.3
(MPa) #
W
0
(mW) —# 10.6 #
min of
[W
.3
(z
1
),I
TA.3
(z
1
)]
(mW)
—
z
1
(cm) —
z
bp
(cm) —
z
sp
(cm) # 0.6
d
eq
(z
sp
) (cm) 0.16
f
c
(MHz) # — # — 6.00 #
Dim of A
aprt
X(cm) —# — 0.16#
Y(cm)
—# — 0.30 #
Other Information
PD (μsec) #
PRF (Hz) #
p
r
@PII
max
(MPa) #
d
eq
@Pll
max
(cm) 0.16
Focal Length FL
x
(cm) —# — #
FL
y
(cm) —# — #
I
PA.3
@MI
max
(W/cm
2
)
#
Operating
Control
Conditions
Control 1: Exam Type Vas, Nrv,
Ven
Control 2: Sample Volume 8 mm
Control 3: PRF 7813
Control 4: Sample Vol. Position Zone 0