26 Imaging modes  
M Mode imaging
Motion mode (M Mode) is an extension of 2D. It 
provides a trace of the 2D image displayed over 
time. A single beam of ultrasound is transmitted, 
and reflected signals are displayed as dots of 
varying intensities, which create lines across the 
screen.
To display the M-line 
1 Press the MMODE key.
Note: If the M-line does not appear, make sure 
that the image isn’t frozen.
2 Use the touchpad to position the M-line where 
desired.
3 Set controls as desired.
Many optimization and depth controls available 
in 2D imaging are also available in M Mode 
imaging. See “2D controls” on page 25.
To display the M Mode trace 
1 Display the M-line.
2 Adjust the depth if necessary. (See “To adjust 
depth” on page 30.)
Brightness Adjusts the display brightness. 
Settings range from 1 to 10.
The display brightness affects 
battery life. To conserve battery life, 
adjust brightness to a lower setting.
Guide Turns guidelines on and off. 
Guidelines are for needle guidance, 
are an optional feature, and depend 
on transducer type.
For transducers with a single-angle 
or multi-angle bracket, the 
touchpad moves the depth cursor.
If the transducer uses a multi-angle 
bracket, select Guide and then 
select the angle: A, B, or C. To exit 
angle selection, select Back. To clear 
the guides, do either of the 
following:
• Select the angle again (A, B, or C).
• Exit angle selection and press 
Guide.
See also the needle guide’s user 
documentation.
Guide is not available when the 
ECG cable is connected.
Sector (Cardiac exam) Specifies the 
sector width. 
SonoMB On is available only for 
Sector Full.
SonoMB
(MB)
MB On and MB Off turn SonoMB® 
multi-beam imaging technology on 
and off. When SonoMB is on, 
MB
appears in the upper left-hand 
screen.
SonoMB depends on transducer 
and exam type. 
MBe See “Needle visualization” on 
page 31.
ECG Displays the ECG trace. See “ECG 
Monitoring” on page 41.
This feature is optional and requires 
a SonoSite ECG cable.
Clips Displays the clip controls. See “To 
capture and save a clip” on 
page 38.
THI  Turns Tissue Harmonic Imaging on 
and off.
When on, 
THI
 appears in the upper 
left-hand screen. This feature 
depends on transducer and exam 
type.
Page x/x Indicates which page of controls is 
displayed. Select to display the next 
page.