User guide
Chapter 3:  Imaging 27
Imaging
3 Press the MMODE key.
The time scale above the trace has small marks 
at 200ms intervals and large marks at 
one-second intervals.
4 Do any of the following as needed:
• Select the sweep speed   (Slow, Med, 
or Fast).
•Press the 
UPDATE key to toggle between the 
M-line and M-Mode trace.
• If using a duplex layout, press the 
MMODE 
key to toggle between the full-screen 
M-line and the duplex layout.
To set a duplex layout, see “Presets 
setup” on page 22.
CPD and color Doppler imaging
Color power Doppler (CPD) and color Doppler 
(Color) are optional features.
CPD is used to visualize the presence of detectable 
blood flow. Color is used to visualize the presence, 
velocity, and direction of blood flow in a wide range 
of flow states.
To display the CPD or Color image
1 Press the COLOR key.
A ROI box appears in the center of the 2D 
image.
2 Select CPD or Color.
The current selection also appears in the upper 
left-hand screen.
The Color indicator bar on the upper left-hand 
screen displays velocity in cm/s in Color imaging 
mode only.
3 Using the touchpad, position or resize the ROI 
box as needed. Press the 
SELECT key to toggle 
between position and size.
While you position or resize the ROI box, a green 
outline shows the change. The ROI box indicator 
on the left-hand screen shows which touchpad 
function is active.
4 Adjust controls as desired. See “CPD and 
Color controls.”
CPD and Color controls
In CPD or Color imaging, you can set the following 
on-screen controls.
Color, CPD Toggle between CPD and Color. 
The current selection appears in 
the upper left-hand screen.
Color 
Suppress
Shows or hides color information. 
You can select Show or Hide 
while in live or frozen imaging. The 
setting shown on-screen is the 
current selection.
Flow 
Sensitivity
The current setting appears 
on-screen.
• Low optimizes the system for 
low flow states.
• Med optimizes the system for 
medium flow states.
• High optimizes the system for 
high flow states.
PRF Scale Select the desired pulse repetition 
frequency (PRF) setting by 
pressing the control keys.
There is a wide range of PRF 
settings for each Flow Sensitivity 
setting (Low, Med, and High).
Available on select transducers.
Wall Filter Settings include Low, Med, and 
High.
Available on select transducers.
Steering Select the steering angle setting of 
the color ROI box (-15, 0, or 
+15). If adding PW Doppler, see 
“PW Doppler controls” on 
page 29.
Available on select transducers.