User guide
Chapter 3:  Imaging 31
Imaging
Needle visualization
About MBe
The MBe control turns on SonoMBe™ imaging, 
which enhances linear structures within a selected 
angle range and can facilitate needle guidance 
during catheter placement and nerve-block 
procedures. A three- or four-sided outline indicates 
the enhancement area. (See Figure 1 on 
page 31.)
For curved array transducers, MBe can help identify 
the direction of the needle, although only segments 
of the needle shaft may show in the image. (See 
Figure 2 on page 32.) Use movement and fluid 
injection to help verify the needle-tip location.
The MBe control is available in full-screen imaging 
only and on the following:
• Transducers: C60x, HFL38x, HFL50x, L25x, 
L38xi
• Exams: Breast, Musculoskeletal, Nerve, Small 
Parts, Vascular (L25x only), Venous (L25x 
only)
Figure 1 Image with MBe on (linear transducer):
1 - Needle
2 - Outlined area enhanced by MBe
3 - Dotted line
4 - Unenhanced area
WARNING:
To avoid incorrect needle 
placement when MBe is on:
• Using movement and fluid 
injection, verify the needle-tip 
location and trajectory. MBe 
enhances linear structures within 
a selected angle range on the 
ultrasound plane. Linear 
structures outside the selected 
angle range or the ultrasound 
plane
—such as a bent 
needle
—may be less apparent.
• Note that linear structures are 
enhanced only in an outlined 
portion of the image. The area 
outside the outline remains 
unchanged. (See Figure 1 on 
page 31.)
• Note that the beam divergence 
of a curved array transducer may 
prevent a segment of the needle 
shaft from showing in the image. 
(See Figure 2 on page 32.) The 
needle tip may not show.
1
2
3
4