Chapter 3: Imaging 41
Imaging
5 Select Export.
The files are finished exporting approximately
five seconds after the USB animation stops.
Removing the USB storage device or turning off
the system while exporting may cause exported
files to be corrupted or incomplete. To stop
in-progress exporting, select Cancel Export.
To delete images and clips
1 In the patient list, do one of the following:
If deleting a single image or clip, display it.
(See “To review images and clips” on
page 40.)
If deleting entire patient exams, select them.
2 Select Delete.
A confirmation screen appears.
To manually archive images and clips
You can send patient exams to a DICOM printer or
archiver, or to a PC using SiteLink Image Manager.
DICOM and SiteLink Image Manager are optional
features. For more information about archiving, see
the DICOM and SiteLink Image Manager
documentation.
1 Select one or more patients in the patient list.
2 Select Archive.
To display information about a patient
exam
1 On the patient list, select the exam.
2 Select Info.
ECG Monitoring
ECG Monitoring is an optional feature and requires
a SonoSite ECG cable.
To monitor ECG
1 Connect the ECG cable to the ECG connector on
the ultrasound system, mini-dock, or docking
system.
ECG Monitoring turns on automatically.
Note: An external ECG monitor may cause a lag
in the timing of the ECG trace, corresponding
with the 2D image. Biopsy guidelines are not
available when ECG is connected. The ECG
signal may take up to one minute to restabilize
after defibrillator use on the patient.
2 Select ECG on-screen. (ECG may be on
another page. It appears only if the ECG cable is
connected.)
3 Adjust controls as desired.
WARNING:
To prevent misdiagnosis, do not
use the ECG trace to diagnose
cardiac rhythms. The SonoSite ECG
control is a non-diagnostic feature.
WARNING:
To avoid electrical interference
with aircraft systems, do not use
the ECG cable on aircraft. Such
interference may have safety
consequences.
Caution:
Use only accessories
recommended by SonoSite with
the system. Your system can be
damaged by connecting an
accessory not recommended by
SonoSite.