Chapter 4:  Measurements and Calculations 49
Measurements
Figure 2 Percent area reduction calculation of right 
carotid bulb 
To calculate percent area reduction
The percent area reduction calculation involves two 
manual trace measurements. 
1 On a frozen 2D image, press the 
CALCS key.
2 Do the following for A
1
 and then for A
2
:
a From the calculations menu, select the 
measurement name under Area Red.
b Using the touchpad, move the caliper to the 
trace starting point, and press the 
SELECT 
key.
c Using the touchpad, trace the desired area. 
To make a correction, select Undo on-screen 
or press the 
BACKSPACE key.
d Complete the trace, and press the 
SET key.
e Save the calculation. See “To save a 
calculation” on page 47.
The percent area reduction result appears 
on-screen in the measurement and calculation 
data area and in the patient report.
To calculate percent diameter reduction
1 On a frozen 2D image, press the CALCS key.
2 Do the following for D
1
 and then for D
2
: 
a From the calculations menu, select the 
measurement name under Dia Red.
b Position the calipers. (See “Working with 
calipers” on page 43.)
c Save the calculation. See “To save a 
calculation” on page 47.
The percent diameter reduction result appears in 
the measurement and calculation data area and in 
the patient report.
Transducer Exam Types
C11x Abdomen, Vascular
C60x Abdomen, Msk
HFL38x IMT, Small Parts, Vascular
HFL50x Msk, Small Parts
L25x Msk, Vascular
L38x IMT, Small Parts, Vascular
L38xi IMT, Msk, Small Parts, Vascular
P10x Abdomen
P21x Abdomen
SLAx Msk, Vascular