Installation manual

18
Chapter3Troubleshooting
Problem Possible cause and remedy
• Themessageshowsthattheinputsignalis
outofthespeciedfrequencyrange.
Example:
fD: Dotclock
fH: Horizontalscanfrequency
fV: Verticalscanfrequency
• CheckwhetherthePCisconguredtomeettheresolution
andverticalscanfrequencyrequirementsofthemonitor
(see“2-1.CompatibleResolutions”(page13)).
• RebootthePC.
• Selectanappropriatesettingusingthegraphicsboard’s
utility.Refertothemanualofthegraphicsboardfor
details.
Error code table
Error Code Description
0*** • ErrorsthatoccurredduringSelfCalibration
1*** • ErrorsthatoccurredduringGrayscaleCheck
*1** • ErrorsthatoccurredduringDICOM
*2** • ErrorsthatoccurredduringCustom
*3** • ErrorsthatoccurredduringCAL1
*4** • ErrorsthatoccurredduringCAL2
*5** • ErrorsthatoccurredduringCAL3
**10 • Theproductmaximumbrightnessmaybelowerthanthetargetbrightness.
• Lowerthetargetbrightness.
**11 • Theproductminimumbrightnessmaybehigherthanthetargetbrightness.
• Raisethetargetbrightness.
**34 • Thesensormaynothavecomeoutduringcalibration,orlightmayhaveenteredthe
sensor.
• TurnofftheACadaptermainpower,waitforafewminutesbeforeturningthepower
backon,andthenexecuteSelfCalibration/GrayscaleCheckagain.
**61 • Thesensormaynothavecomeout.
• Checkwhetherthereisanyforeignobjectnearthesensor.
• ExecuteSelfCalibration/GrayscaleCheckagain.
**94 • SelfCalibration/GrayscaleCheckcannotbeexecutedduetothefollowingsettings:
- Thereisnosignalinputtotheleftscreen,andtheverticalresolutionoftheright
screenislowerthan2048dots.
- Theverticalresolutionsoftheleftandrightscreensarelowerthan2048dots.
**95 • TheexecutionconditionsforSelfCalibration/GrayscaleCheckmaynotbeappropriate.
• Checktheexecutionconditionsandresetthemifnecessary.