Installation manual

27
Appendix
English
EMC Information
Essential performance of RadiForce series is to display images and operate functions normally.
CAUTION
The RadiForce series requires special precautions regarding EMC and need to be installed, put into
service and used according to the following information.
Do not use any cables other than the cables that provided or specied by us.
Using other cables may cause the increase of emission or decrease of immunity.
Cable Length: Max. 3m
Do not put any portable and mobile RF communications equipment close to the RadiForce series.
Doing so may affect the RadiForce series.
The RadiForce series should not be used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the equipment or system should be observed to verify normal
operation in the conguration in which it will be used.
Anyone who connects additional equipment to the signal input part or signal output parts,
conguring a medical system, is responsible that the system complies with the requirements of
IEC/EN60601-1-2.
Guidance and manufacturer’s declaration - electromagnetic emissions
TheRadiForceseriesisintendedforuseintheelectromagneticenvironmentspeciedbelow.
ThecustomerortheuseroftheRadiForceseriesshouldassurethatitisusedinsuchanenvironment.
Emission test Compliance Electromagnetic environment - guidance
RFemissions
CISPR11/EN55011
Group1 TheRadiForceseriesusesRFenergyonlyforitsinternalfunction.
Therefore,itsRFemissionareverylowandarenotlikelytocauseany
interferenceinnearbyelectronicequipment.
RFemissions
CISPR11/EN55011
ClassB TheRadiForceseriesissuitableforuseinallestablishments,includingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
Harmonicemissions
IEC/EN61000-3-2
ClassD
Voltageuctuations/
ickeremissions
IEC/EN61000-3-3
Complies
Guidance and manufacturer’s declaration - electromagnetic immunity
TheRadiForceseriesisintendedforuseintheelectromagneticenvironmentspeciedbelow.
ThecustomerortheuseroftheRadiForceseriesshouldassurethatitisusedinsuchanenvironment.
Immunity test IEC/EN60601 test
level
Compliance level Electromagnetic environment - guidance
Electrostatic
discharge(ESD)
IEC/EN61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floorsshouldbewood,concreteorceramictile.
Ifoorsarecoveredwithsyntheticmaterial,the
relativehumidityshouldbeatleast30%.
Electricalfast
transient/burst
IEC/EN61000-4-4
±2kV
forpowersupplylines
±1kV
forinput/outputlines
±2kV
forpowersupplylines
±1kV
forinput/outputlines
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Surge
IEC/EN61000-4-5
±1kVline(s)toline(s)
±2kVline(s)toearth
±1kVline(s)toline(s)
±2kVline(s)toearth
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Voltagedips,short
interruptionsand
voltagevariationson
powersupplyinput
lines
IEC/EN61000-4-11
<5%U
T
(>95%dipin
U
T
)for0.5cycle
40%U
T
(60%dipinU
T
)
for5cycles
70%U
T
(30%dipinU
T
)
for25cycles
<5%U
T
(>95%dipin
U
T
)for5sec
<5%U
T
(>95%dipin
U
T
)for0.5cycle
40%U
T
(60%dipin
U
T
)for5cycles
70%U
T
(30%dipin
U
T
)for25cycles
<5%U
T
(>95%dipin
U
T
)for5sec
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.Iftheuser
oftheRadiForceseriesrequirescontinued
operationduringpowermainsinterruptions,it
isrecommendedthattheRadiForceseriesbe
poweredfromanuninterruptiblepowersupplyor
abattery.
Powerfrequency
(50/60Hz)
magneticeld
IEC/EN61000-4-8
3A/m 3A/m Powerfrequencymagneticeldsshouldbeat
levelscharacteristicofatypicallocationina
typicalcommercialorhospitalenvironment.
NOTEU
T
isthea.c.mainsvoltagepriortoapplicationofthetestlevel.