Troubleshooting
PROBLEM POSSIBLECAUSE SOLUTION
LCDdisplay
showsabnormal
result
Thecupositionisnotcorrect
orcuisnotproperlytightened
Applythecucorrectly
andtryagain
Bodyposturewasnotcorrect
duringtesting
Reviewthe“BODYPOSTURE
DURINGMEASUREMENT”
sectionsoftheinstructions
andre-test.
Talking,armorbodymovement,
irritation,excitementortension
duringtesting.
Re-testwhencalmandwithout
speakingormovingduring
thetest
Irregularheartbeat(arrhythmia) Itisinappropriatefor
peoplewithseriousarr-
hythmiatousethisblood
pressuremonitor.
Display messages in LCD
Lowbattery Lowbattery Changebatteries
symbol
Er0 Pressuresystemisunstable Don’tmoveandtryagain.
beforemeasurement
Er1 Failtodetectsystolicpressure
Er2 Failtodetectdiastolicpressure
Er3 Pneumaticsystemblockedorcu Applythecucorrectly
istootightduringination andtryagain
Er4 Pneumaticsystemleakageorcu
istoolooseduringination
Er5 Cupressureabove300mmHg Wait5minutesand
measureagain.Ifthemonitor
 isstillabnormal,please
contactthelocaldistributor
orthemanufatcurer.
Er6 Morethan3minuteswithcu
pressureabove15mmHg
Er7 EEPROMaccessingerror
Er8 Deviceparametercheckingerror
Er9 MCUself-verifyerror
ErA Pressuresensorparametererror
Noresponse Incorrectoperationorstrong Takeoutbatteriesand
whenyoupress electromagneticinterference. andreinstallafter5minutes.
buttonorload
batteries.