7.2 Switching to Auto-Flex
Important: Before proceeding with Auto-Flexion, read Section 3.6 on ‘Basic
Flexion (Balancing the Patient)’.
It is very important that you become familiar with how to balance your patient
BEFORE USING the Auto-Flex feature. THE PATIENT MUST BE SUPERVISED AT ALL
TIMES DURING OPERATION.
1. Balance your patient by following steps 1 - 4 under section 3.6 “Basic
Flexion”. Patient should start in a slightly flexed position.
2. To begin using the Auto-Flexion Touch
Screen, press the “FLEX” button
(Photo 21) on the main control panel.
The touch screen should already be
lit. By default, the touchscreen displays
the settings from the last treatment.
At start up, the touchscreen displays
the Auto-Distraction screen (if so
equipped). Press the “Dist/Flex” button
to switch to the Auto-Flexion display.
3. Set each of your touchscreen settings
by touching each button in turn
(buttons will turn blue when selected)
and adjusting the slider for the de-
sired setting.
Note: Adjusting the slider works best
with the back of your finger or by sim-
ply tapping the slide bar at the desired point on the slide bar.
4. With settings selected, press “RUN” to start the treatment and move the
flexion locking handle to the 1 o’clock position to engage the motor (Photo
22 - see note below). You will feel the handle move a cam into a lock position.
Note: Once the RUN Button is pressed, the operator has 30 seconds to engage the
Auto-Flex motor by turning the flexion lock handle to the 1 o’clock position (photo 22).
If not engaged within that time, the motor will shut off.
6. Press the yellow “PAUSE” button at any time to pause the treatment or to
adjust your settings. Press PAUSE again to restart the treatment.
7. When the cycle time ends, the table will return to the rest/horizontal
position. Before the patient exits the table, be sure the flexion, rotation
and lateral lock handles are all turned to the 6 o’clock/locked position.
Note: When using distraction with Auto-Flex, flexion may not need to be as
deep. Making a note of your settings for each patient makes it easy to dupli-
cate treatment on subsequent visits.
photo 21
photo 22