Emergency Stopping for Auto-Distraction
To immediately stop an Auto-Distraction treatment at any time, press the
STOP button on the
touchscreen (right). The
patient can also stop
the treatment with the
Stop-Treatment Patient
Switch, shown right.
Instructions for the Auto-Distraction Option
To use distraction on the Air-Flex, first
press the “DIST” button on the main con-
trol panel (photo 23). If this is the first
treatment after table start-up, the touch-
screen will automatically display the
Auto-Distraction display. If you have al-
ready been using the touchscreen for
Auto-Flexion, simply press the “Dist/Flex”
button to enter Auto-Distraction.
NOTE: The Air-Flex can be locked in any flexion or lateral position during
Auto-Distraction.
8.1The Auto-Distraction Touchscreen
Auto-Distraction functions defined:
REST - Describes the duration of the
treatment Rest Cycle in seconds. lt is
programmable from 5 to 60 seconds.
During a treatment, the Rest Cycle will
highlight and countdown the seconds
remaining in that cycle.
PULL - Describes the duration of the treatment Pull Cycle in seconds. lt is
programmable from 5 seconds to 60 seconds. During a treatment, the Pull
Cycle will highlight and countdown the seconds remaining in that cycle.
REST% - Rest % is used to calculate the Force during the Rest Cycle. lt can
be set to either 25% or 50%. For example, with a Force Setting of 40 lbs
during the Pull Cycle, setting the Rest % to 25% will administer 10 lbs of
Force during the Rest Cycle. A 50% setting yields 20 lbs. during Rest Cycle.
TIME - Describes the duration of the treatment time in minutes. The display
will count down once the treatment has begun.
FORCE - The treatment Force is measured in U.S. pounds. It is programmable
from 5 lbs-100 lbs for a Lumbar Treatment and 2 lbs–50 lbs for a Cervical
photo 23
Patient