Treatment. During a treatment, the Force Setting will show the measured
force in real-time.
ARROWS - The Up and Down Arrow Buttons are used to increase and de-
crease the treatment setting values. To edit a value, simply touch it and it
will highlight in BLUE. Then use the arrow keys to enter the desired value.
LUMBAR/CERVICAL - The CERV/LMBR button indicates which distraction
treatment type the table is currently set for.
If your table is equipped for both lumbar
and cervical distraction, pressing the
CERV/LMBR button will toggle between the
two functions respectively. If your Air-Flex
is only equipped with one auto-distraction
option, you will not need to use this button.
MAN/AUTO - Manual/Automatic Button in-
dicates which type of distraction mode
(see below) you want to use. Press the button to toggle back and forth be-
tween MAN and AUTO.
RUN: Pressing RUN starts the treatment. While the system is in run mode
the status display will flash "Run Mode Active" (see below).
8.2 Manual/Auto Mode
Manual Mode (MAN)
Manual Mode allows you to manually distract your patient using the pressure
foot pedal. The Force Setting shows the pounds being administered to
your patient once you release the foot pedal. This force will now be used
as the peak force once the RUN button is pressed.
NOTE: For safety, Manual mode is not available while in Cervical Distraction.
Automatic Mode (AUTO)
Automatic Mode allows the table to automatically distract your patient with
the amount of force you have set. The lumbar default setting is 50 lbs.; the
cervical default setting is 10 lbs. Settings can also be changed before or
during treatment using the up and down arrows.
Run Mode
Once the RUN button is pressed, the treatment
begins and the status display flashes “Run Mode
Active”. While in Run Mode, the system cycles from
Pull to Rest while displaying the actual Force being
distributed in real-time in the FORCE display.