User's Manual Part 3

121
The calibration procedure of the module requires a complete structure of display board (see MTG0079)
in order to perform the software selection which will be referred to later and power the module itself..
- Menu of the External Referement Module
Verification of the external reference section  connect a spectrum analyser to the BNC labelled EXT
REF IN/OUT and check the sections therein:
q Configure the module with ExtREF Dir Out and check the presence of the 5MHz carrier with
ExtReference set to Pres in the display menu.
q Check that the 5MHz frequency is fine-adjusted by acting on ExtRef Tune.
Verification of the IF Link section:
q Configure the module with IF Link open and closed checking the switching of the relays.