Operating instructions
Operation SW 5250A•SW 3500A•SW 1750A
3-24 Operation Manual
SYNC OUT
selects the output sync signal to ON, OFF or EVENT. In the
ON
position,
the signal is driven by the output frequency of the waveform, or sequence settings. In
the
EVENT
position, a pulse will be generated for any change in VOLT, CURR, FREQ,
PHASE or WAVEFORM.
The
SYNC OUT
setting only affects normal (non-sequence) mode. To set up the SW
so that a sync pulse only occurs for a specified sequence segment, set this parameter
to OFF. That way, before and after a sequence is run, no other sync pulses will occur.
Table 3–2 provides self-test result definitions.
0001 LCD Test 0100 Battery Test
0002 EPROM CKSUM 0200 QSPI Test
0004 NVRAM BANK 0 CKSUM 0400 DFI Test
0008 NVRAM BANK 1 CKSUM 0800 Speaker Test
0010 RAM Test 1000 DWSB Self-Test 0
0020 VXI Test 2000 DWSB Self-Test 1
0040 QSCI Test 4000 T&MB Self-Test 0
0080 GPIB Test 8000 T&MB Self-Test 1
Table 3–2 SELF-TEST Results Definitions