Operating instructions
Operation SW 5250A•SW 3500A•SW 1750A
3-34 Operation Manual
LOAD
EDIT
VIEW
SAVE
DELETE
FREQ Hz
Vrms V
START Deg
TIME ms
STOP Deg
AMPL V
60.00
120.00
0.00
4.17
90.00
0.00
WAVEFORM OUTPUT [OFF]
9. Use either the Knob or the Up/Down Arrow keys to select
FREQ
. The frequency
is normally defaulted at 60 Hz with all other values set to 0. On the keypad, enter
6
,
0
,
decimal point
,
0
,
0
, then press <
Enter
> to set the frequency to 60.00 Hz.
10. Use either the Knob or the Up/Down Arrow keys to select
Vrms
. Using the
keypad, enter
1
,
2
,
0
, then press <
Enter
> to set the RMS voltage to 120 volts.
11. Use either the Knob or the Up/Down Arrow keys to select
START
. Using the
keypad, enter
0
, then press <
Enter
> to set the start phase angle to 0°.
12. Although the TIME can be calculated then entered, the STOP function is used in
this example. Select
STOP
, enter
90
for a 90° phase angle, and press <
Enter
>.
Note that the unit automatically calculates the
TIME
(in this case, 4.17 ms).
13. Since the operator is not programming a transient (drop/spike), leave the
AMPL
(amplitude) at 0V.
14. Press the
Menu
key or the
Left Arrow
key to return to the WAVEFORM Menu.
LOAD
EDIT
VIEW
SAVE
DELETE
WAVEFORM OUTPUT [OFF]