18. While still in the PHASE A sub-menu:
a. Select
AMPL
(amplitude) with the Knob or Arrow Keys.
b. Enter
120
with the keypad to set the amplitude at 120 volts, then press
<
Enter
>. The inrush waveform will now be displayed on Phase A output on
the front panel.
19. While still in the PROGRAM mode:
a. Select
PHASE B
by cursoring to Phase A, then press <
Enter
>.
b. Select
FUNC
with the Knob, then press <
Enter
> to place a box around the
waveshape selection.
c. Rotate the Knob to select
TRIANGLE
, then press <
Enter
>.
d. Select
AMPL
(amplitude) with the Knob, enter
120
with the keypad, then
press <
Enter
>. A triangular waveform will be displayed on Phase B output
on the front panel.
e. With the field select cursor still at
AMPL
, press <
Enter
> to place a box
around the waveshape selection then, using the Knob, vary the amplitude
of the triangular waveform on Phase B.
f. Select
PHASE C
by cursoring to Phase B, then press <
Enter
>.
g. Select
FUNC
with the Knob, then press <
Enter
> to place a box around the
waveshape selection.
h. Rotate the Knob to select
SQUARE
, then press <
Enter
>.
i. Select
AMPL
(amplitude) with the Knob, enter
120
with the keypad, then
press <
Enter
>. A square waveform will now be displayed on Phase C
output on the front panel.
j. With the field select cursor still at
AMPL
, press <
Enter
> to place a box
around the amplitude selection then, using the Knob, vary the amplitude of
the square waveform on Phase C.
20. Press <
Menu
> or the
Left Arrow
key to return to the MAIN MENU.