Operating instructions

Operation SW 5250A•SW 3500A•SW 1750A
3-40 Operation Manual
15. To retain the waveshape and display it on Phase A output on the front panel:
a. Return to the MAIN MENU by pressing <
Menu
> twice.
b. Select the
PROGRAM
Menu, then press <
Enter
>.
PROGRAM
MEASURE
WAVEFORM
SEQUENCE
INSTR
SYSTEM
PHASE A
AMPL 120.00 V
CURL 5.00 A
FREQ 60.00 Hz
ANG 0.00 Deg
FUNC SCRATCH
MAIN MENU OUTPUT [OFF]
θ
16. In the PHASE
A sub-menu:
a. Select
FUNC
with the Knob, then press <
Enter
> to place a box around the
waveshape selection.
b. Rotate the Knob until
SCRATCH
is displayed, then press <
Enter
>. The
scratchpad waveform will now be displayed on Phase A of the oscilloscope.
17. Press <
Menu
> or the
Left Arrow
key to return to the MAIN MENU.