Operating instructions

SW 5250A•SW 3500A•SW 1750A Standard Waveforms
Operation Manual 4-3
  


D

Spike250 Sine with spikes at peaks. Spikes are from 85 to 95°
and calibrated to be 250Vpeak when programmed to
120Vrms
1.4731 40-5000 Hz
Spike300 Sine with spikes at peaks. Spikes are from 85 to 95°
and calibrated to be 300Vpeak when programmed to
120Vrms
1.7678 40-5000 Hz
Spike400 Sine with spikes at peaks. Spikes are from 85 to 95°
and calibrated to be 400Vpeak when programmed to
120Vrms
2.3570 40-5000 Hz
Square Square wave with 50% duty cycle (true rms) 0.7071 40-5000 Hz
Step_06 Six voltage step sine wave 1.0 40-5000 Hz
Step_12 Twelve voltage step sine wave 1.0 40-5000 Hz
Step_24 Twenty-four voltage step sine wave 1.0 40-5000 Hz
Taylor 5th harmonic Taylor series wave (true rms) 0.9136 40-5000 Hz
Triangle Triangle wave (true rms) 1.2246 40-5000 Hz
Z1 3-phase output with a voltage spike occurring at the same
point in time. Spikes are from 45 to 90° and calibrated to
be 200 Vpeak when programmed to 80 Vrms
1.0 40-5000 Hz
Z2 3-phase output with a voltage spike occurring at the same
point in time. Spikes are from 165 to 210° and calibrated
to be 200 Vpeak when programmed to 80 Vrms
1.0 40-5000 Hz
Z3 3-phase output with a voltage spike occurring at the same
point in time. Spikes are from 285 to 330° and calibrated
to be 200 Vpeak when programmed to 80 Vrms
1.0 40-5000 Hz