User's Manual

Table Of Contents
TrackSense
®
LyoPro Version 3.3 Page 14 of 53
User Manual User Manual TrackSense LyoPro MAN 7
Note: In the event of a failure due to ESD, the Access Point will obtain a
permanently reduced sensitivity.
3.3.3. Communication SetupLIVE Hardware Settings
Alias: insert name of template
Select Transmission Rate:
5-9 sec for up to 32 loggers
10-19 for 32-64 loggers
20 or higher for 100 loggers
Important: Please note that due to limitations of the data package size, the ratio
between sample rate and transmission rate is automatically set to be less than
1:20.
Select channel numbers for antenna A and B or use suggested default values.
3.3.3.1. USB
1. Connect the USB cable to the Access Point and PC
2. Verify that the USB is providing power to the Access Point (power
light should be green)
3. Once a LIVE data session has been started, verify that the Wireless
LED light on the Access Point is blinking green, indicating that
communication between the logger and Access Point has been
established
4. Insert name and choose connection mode: USB
5. Once the access Point has been defined and set up, drag and drop
the box into the Template box: