User's Manual

Table Of Contents
TrackSense
®
LyoPro Version 3.3 Page 15 of 53
User Manual User Manual TrackSense LyoPro MAN 7
6. The completed template can now be selected and used when
starting a session
3.3.3.2. LAN
Note: When setting up an IP address on the Access Point, a USB cable must be
connected between the Access Point and PC.
1. Connect the Ethernet cable to the Access Point and LAN network
2. Connect the 5VDC power supply and USB cable to the Access Point.
Reference the software manual in order to properly setup the IP
address in the ValSuite software
3. Once a LIVE session has started, verify that the Wireless LED light on
the Access Point is blinking green, i.e. indicating that communication
between the data logger and Access Point has been established
4. Insert name and choose connection mode: Ethernet (Static or DHCP)