LyoPro Version 3.3 Page 20 of 53
User Manual User Manual TrackSense LyoPro MAN 7
1.
Ensure that the battery has enough capacity to cover the expected session
length. If the battery is new, activate it separately in the delivered activator for
60 minutes before placing it in the LyoPro data logger with the
+
pole facing
downwards
2.
If not already mounted, connect the thermocouple sensor to the data logger
using the supplied screwdriver. It is important that the RED wire goes in the
RED screw terminal. And that the BLUE wire goes in the BLUE terminal. Do not
tighten too hard in order to avoid possible damage to the thermocouple
3.
Place the data logger(s) in the reader station slot(s)
4.
Select LyoPro
Hardware Settings in ValSuite main menu and chose USB
(default) or Ethernet connection and click OK