Reading TrackSense LyoPro Data Loggers and Creating Reports:
12.
Place the data loggers into the Reader Station. Go to LyoPro
Read Logger,
then Select All and Scan
13.
Select Read. The software will read all data loggers and display the data in the
graph. Select Close
14.
From the main menu, choose Session Add to select the appropriate Reports
and Data Analyzes
15.
To print a report, click File
Print. A Print window will appear with various
options. Select the desired modules and click OK
Graph Features Review:
Right click the mouse when it is placed on the graph:
A. Select Data Viewercompare data sets
B. Select Interval Data Viewernumerically display data
C. Select Add Labelplace special information directly in the graph
D. Select Copy Graphcopy graph to the clipboard
E. Select Set Axis Min/Maxscale a graph either automatically or fixed
F. Select Propertiesdefine data series
For more information on the many other ValSuite software features, please consult
the separate manual delivered with the system.
6. TrackSense
®
LyoPro Loggers
6.1. TrackSense
®
LyoPro
6.1.1. Data Logger Specifications
Ordering
codes
Logger type Temp Range
Operating
Pressure
Data
Points
Dimensions
Sensor
31201000
Measuring Range:
-62 to +62 °C
Calibrated Range:
-60 to 60 °C
0.001 mBar
to 4 Bar abs.
100,000
2x
Ø18x32mm
Min. 1 sec.
Max. 23:59
(Hours: Min)
Thermocouple
mounted by
screw
terminal