User's Manual

Table Of Contents
TrackSense
®
LyoPro Version BETA Page 15 of 46
User Manual User Manual TrackSense LyoPro MAN 7
4.1.2. Releasing the License
Users who wish to release the license so that it may be used on another PC,
must follow the procedure below:
1. Select Help About ValSuite
2. Click License
3. Click Clear License. This will remove the license from this PC. The
license will then be released and ready to be used again on another
PC
5. How to Start Logging Data with TrackSense LyoPro
Starting TrackSense LyoPro Loggers:
1.
Ensure that the battery has enough capacity to cover the expected session
length. If the battery is new, activate it separately in the delivered activator for
60 minutes before placing it in the LyoPro data logger with the
+
pole facing
downwards
2.
If not already mounted, connect the thermocouple sensor to the data logger
using the supplied screwdriver. It is important that the RED wire goes in the
RED screw terminal. And that the BLUE wire goes in the BLUE terminal. Do not
tighten too hard in order to avoid possible damage to the thermocouple