User's Manual
Table Of Contents
- 1. Safety Information
- 2. New Order Contents
- 3. TrackSense® LyoPro 10-slot Reader Station
- 4. Activating ValSuite®
- 5. How to Start Logging Data with TrackSense LyoPro
- 6. TrackSense® LyoPro Loggers
- 6.1. TrackSense® LyoPro
- 6.1.1. Data Logger Specifications
- 6.1.2. Proper Use
- 6.1.3. Data Logger Serial Numbers
- 6.1.4. Disassembling / Assembling Data Loggers
- 6.1.5. Procedure for Cleaning TrackSense® LyoPro Data Loggers
- 6.1.6. Procedure for Changing the Data Logger O-Ring
- 6.1.7. Extended Warranty
- 6.1.8. Thermocouple Sensor
- 6.1.9. -30 to -80 C Application Recommendations
- 6.1. TrackSense® LyoPro
- 7. Battery
- 8. Calibration
- 9. Accessories
- 10. Examples on how to Configure LyoPro with Appropriate Accessories
- 11. Firmware Update
- 12. Ellab Service Contract
- 13. Technical Specifications
- 14. FAQ
- 15. Changelog
TrackSense
®
LyoPro Version BETA Page 17 of 46
User Manual User Manual TrackSense LyoPro – MAN 7
7.
If an immediate start (unprogrammed) is required, press Start
8.
The slots will turn green and begin showing the first live data. The data loggers
are now active and can be placed in the process. Press Close
9.
Alternatively, the data loggers can be programmed: Choose Settings at the
bottom of the screen. The Logger Settings window will appear. Define Sample
Rate and optionally Programmed Start/Stop conditions and activate Primary
Control. Choose Select All
Apply
Close
10.
After the settings are setup, select Start. When all selected slots turn green the
data loggers are programmed and will start logging data. Select Close
11.
Go to LyoPro
Logger List. This will showcase all active/programmed data
loggers. Review the details to confirm that they were programmed correctly